Rudolph Research Analytical - AUTOPOL V
The AUTOPOL V Automatic Polarimeter is designed for today's FDA regulated analytical laboratories. Whether you are a...
Active Questions & AnswersAsk a Question
There are no current Discussions
Polarimeters Service ProvidersView All (3)
Documents & ManualsView All Documents
Features of AUTOPOL V
- TempTrol™ - Electronic cooling and heating from 15°-30°C
- 21CFR Part 11 Compliance: Electronic signature and secure local data storage
- Six Standard Wavelengths: 365nm, 405nm, 436nm, 546nm, 589nm, 633nm
- Standard Accessories: TempTrol™ NIST Traceable Quartz Standard, TempTrol™ 100mm Polarimeter Cell, TempTrol™ Temperature Validation Cell and Built-In Sample Measurement Probe.
General Specifications
Depth | 17 in |
Height | 10.5 in |
Width | 35 in |
Weight | 80 lbs (operational) |
Additional Specifications
Measuring Mode:
Optical Rotation
Specific Rotation
Concentration
User Defined Scale
Measuring Scale: Degrees Arc, % Concentration
Measuring Range:
± 89° Arc Optical Rotation
± 999.99° Arc Specific Rotation
0-99.9% Concentration
Resolution:
0.001° Arc Optical Rotation
0.001% Concentration
0.001° Specific Rotation
Reproducibility : 0.002° Arc Optical Rotation
Standard Model Specs:
Accuracy: 0.002° up to 1°, 0.2% up to 5°, 0.01° above 5°
AP Model Specs Accuracy (546nm and 589nm): ±0.002° Arc Optical Rotation
Accuracy for other wavelengths is the same as the standard model
Prism: Glan Thomson calcite quartz
Optical Wavelengths: 365nm, 405nm, 436nm, 546nm, 589nm, 633nm
(other wavelengths available)
Wavelength Selection: Automatic by push-button
TempTrol™ Range: Automatic Electronic Heating & Cooling 15°-35°C
TempTrol™ Accuracy: ± 0.2°C
Measurement Time: 1°/sec slewing rate & 2 sec nominal settling time
Light Source: Tungsten-halogen 6V, 20W, avg. 2,000 hr life
Sample Chamber: Accepts sample tubes up to 200mm
Data Storage: Multimedia Card
Communication Interface: Two RS 232 serial ports, one parallel printer port and one auxiliary port
Analog Output: 0.1% resolution 0-10 volt full range (optional)
Calibration: Automatic calibration by push-button
Display: Graphics LCD, 320 x 240 dots cold fluorescent back lit
User Interface: Touchscreen
Automatic Sensitivity Control:
Permits measurement for sample transmittance as low as 0.01 % or up to O.D. 4.0
Input Power:
100 V, 50/60 Hz
115 V, 50/60 Hz
220 V, 50/60 Hz
Shipping Dimensions:
41"W x 25"H x 25"D
1025mm W x 625mm H x 625mm D
Shipping Weight: 115 lbs (51 kg)